Description
Features
- Measure CP% in situ, near to product.
- Meets CQI-9 daily check requirement.
- Ethernet, USB ports for connection to SSi CATView software for remote management.
- Longlife battery.
- Downloadable data for load traceability.
The CAT100 instantly calculates the Carbon Potential (CP%) by measuring the electrical resistance of a steel coil after it has been soaked in the furnace atmosphere for 30-40 minutes.
A simple routine is followed for easy compliance to CQI-9 daily check requirement: Insert the coil into the furnace, allow for adequate soaking and install on the CAT100. After 30 seconds the CP% is displayed.
A robust sample rod, a portable carry-case, and 100 coils are included with the unit.
Features
| Processes and Applications | Carburising, Hardening |
|---|
| Input Power Range: | Universal, 85 – 264VAC |
| Accuracy: | ± .03% Carbon |
| Read Time: | 30 seconds per coil |
| Measurement Range: | .1 to 1.3% Carbon |
| Operating Temperature: | 50 to 110 °F (10 to 43 °C) |
Datasheet
Manual
Technical Paper